EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #346 background imageLoading...
Page #346 background image
5-264 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:FORCe:STATus
:PAGE:STRess:FORCe:STATus
This command resets the stress status (time and percent) to zero.
Query returns the stress time forced and percent completion of the stress duration
setting.
At *RST, stress time and percent are 0.
Syntax :PAGE:STRess:FORCe:STATus RESet
Parameter
Query response time, rate <newline><^END>
time and rate are NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:FORC:STAT RES"
OUTPUT @Hp4155;":PAGE:STR:FORC:STAT?"
ENTER @Hp4155;A,B
:PAGE:STRess:SETup:ACCumulate
This command sets the amount of stress that has already been forced to the device.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:ACCumulate RESet | accumulate
Parameter
Query response accumulate <newline><^END>
accumulate is NR3 response data type.
Parameter Type Explanation
RESet character reset the status to zero
Parameter Type Explanation
RESet character reset accumulated time to zero
accumulate numeric 0 and above

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals