Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-99
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:VSU<n>:VNAMe
:PAGE:CHANnels[:CDEFinition]:VSU<n>:VNAMe
This command sets the VNAME of VSU<n>.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
At *RST, this value is VSU<n>.
Syntax :PAGE:CHANnels [ :CDEFinition] :VSU<n>:VNAMe name
Parameter
Query response name <newline><^END>
name is the string response data, but does not contain double quote characters at the
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:CHAN:VSU1:VNAME 'VD'"
OUTPUT @Hp4155;":PAGE:CHAN:VSU1:VNAME?"
ENTER @Hp4155;A$
:PAGE:CHANnels:MATrix:CONNection
This command sends the matrix connection setup to the Agilent E5250A Low
Leakage Switch Mainframe.
Execution
Conditions
This command is executable only on the built-in IBASIC controller.
Syntax :PAGE:CHANnels:MATrix:CONNection
Example OUTPUT @Hp4155;":PAGE:CHAN:MAT:GPIB:ADDRESS 20"
OUTPUT @Hp4155;":PAGE:CHAN:MAT:CONTROL ON"
OUTPUT @Hp4155;":MMEM:DEST INTERNAL"
OUTPUT @Hp4155;":MMEM:LOAD:STAT 0,'SET1.MAT','DISK'"
DISP "MODIFY E5250A SETUP. THEN PRESS Continue."
PAUSE
DISP ""
OUTPUT @Hp4155;":PAGE:CHAN:MAT:CONN"
Parameter Type Explanation
name string String of up to 6 alphanumeric characters.
1 st character must be alphabet.