5-8 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:CALibration:SELected?
:CALibration:SELected?
This command performs the specified calibration item (test_number) of the
instrument, then returns a number that indicates the calibration result.
A zero is returned if calibration is completed successfully. Otherwise, a non-zero
value (error number) is returned.
In addition to the error reported by this query response, the 4155C/4156C still
reports calibration errors through the normal status-reporting mechanism.
This command has query form only.
Syntax :CALibration:SELected? test_number
Parameter
Query response result <newline><^END>
Example OUTPUT @Hp4155;":CAL:SEL? 101"
ENTER @Hp4155;A
Parameter Type Explanation
test_number numeric item number of the calibration
100 : VSU1,2/VMU1,2
101 : SMU1
102 : SMU2
103 : SMU3
104 : SMU4
105 : SMU5
106 : SMU6
107 : PGU1,2
108 : GNDU
109 : ADC
Result Explanation
0PASS
1FAIL
2 DONE
3 NOT DONE

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals