5-234 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR2:POINts
:PAGE:MEASure[:SWEep]:VAR2:POINts
This command sets the number of sweep steps of VAR2.
You use this command only if there is an SMU or VSU whose function (FCTN) is
VA R 2 .
At *RST, this value is 5.
Syntax :PAGE:MEASure [ :SWEep] :VAR2:POINts no_of_steps | MINimum | MAXimum
Parameter
Query response no_of_steps <newline><^END>
no_of_steps is NR1 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR2:POINTS 5"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:POINTS MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:POINTS?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:POINTS? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR2:STARt
This command sets the sweep START value of VAR2.
You use this command only if there is an SMU or VSU whose function (FCTN) is
VA R 2 .
At *RST, this value is 20 mA.
Syntax :PAGE:MEASure [ :SWEep] :VAR2:STARt start | MINimum | MAXimum
Parameter
Parameter Type Explanation
no_of_steps numeric 1 to 128
Parameter Type Explanation
start numeric -200 to 200 V or -1 to 1 A. The range of this value
depends on the unit type of VAR2.