5-86 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:MODE
:PAGE:CHANnels[:CDEFinition]:MODE
This command sets the MEASUREMENT MODE. This command also has a query
form. At *RST, this value is set to SWEep.
Syntax :PAGE:CHANnels [ :CDEFinition] :MODE SWEep | SAMPling | QSCV
:PAGE:CHANnels [ :CDEFinition] :MODE?
Parameter
Query response SWE | SAMP | QSCV <newline><^END>
Example OUTPUT @Hp4155;":PAGE:CHAN:MODE SWEEP"
OUTPUT @Hp4155;":PAGE:CHAN:MODE?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition]:PGU<n>:DISable
This command deletes the settings of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
This command does not have query form.
Syntax :PAGE:CHANnels [ :CDEFinition] :PGU<n>:DISable
Example OUTPUT @Hp4155;":PAGE:CHAN:PGU1:DIS"
:PAGE:CHANnels[:CDEFinition]:PGU<n>:MODE
This command sets the output MODE of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
Parameter Type Explanation
SWEep character Sweep measurement mode
SAMPling character Sampling measurement mode
QSCV character Quasi-static CV measurement mode