EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #235 background imageLoading...
Page #235 background image
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-153
SCPI Commands
:PAGE:KSWeep[:MENU]
:PAGE:KSWeep[:MENU]
This command changes the present display page to KNOB SWEEP page.
This command does not have query form.
Syntax :PAGE:KSWeep [ :MENU]
Example OUTPUT @Hp4155;":PAGE:KSW"
:PAGE:KSWeep:SCOpy
This command copies the sweep and graphic display settings of KNOB SWEEP
page to the corresponding parameters of normal sweep.
This command does not have query form.
Syntax :PAGE:KSWeep:SCOPy
Example OUTPUT @Hp4155;":PAGE:KSW:SCOP"
:PAGE:KSWeep:VAR1:EXTent
This command adjusts the sweep extent of the knob sweep measurement. This
parameter is multiplied times the VAR1 range setting to determine the knob sweep
range. For example, if the parameter of this command is set to 0.7, and the VAR1
range is set to 2 V, then VAR1 is swept from 0 to 1.4 V.
At *RST, this value is 0.
Syntax :PAGE:KSWeep:VAR1:EXTent extent
Parameter
Negative value is not allowed. Output polarity is determined by the
:PAGE:KSWeep:VAR1:POLarity command.
Query response extent <newline><^END>
extent is NR3 response data type.
Parameter Type Explanation
extent numeric 0 to 1

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals