Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-85
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:GNDU:VNAMe
:PAGE:CHANnels[:CDEFinition]:GNDU:VNAMe
This command sets the VNAME of GNDU.
At *RST, this value is not defined.
Syntax :PAGE:CHANnels [ :CDEFinition] :GNDU:VNAMe name
Parameter
Query response name <newline><^END>
name is string response data, but does not contain double quote characters at the
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:CHAN:GNDU:VNAME 'VD'"
OUTPUT @Hp4155;":PAGE:CHAN:GNDU:VNAME?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition][:MENU]
This command changes the present display page to "CHANNELS: CHANNEL
DEFINITION" page.
This command does not have query form.
Syntax :PAGE:CHANnels [ :CDEFinition] [ :MENU]
Example OUTPUT @Hp4155;":PAGE:CHAN"
Parameter Type Explanation
name string String of up to 6 alphanumeric characters.
1st character must be alphabet.