Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-201
SCPI Commands
:PAGE:MEASure:QSCV:VAR1:COMPliance
:PAGE:MEASure:QSCV:VAR1:COMPliance
This command sets the compliance value of the VAR1 channel for Quasi-static CV
measurements. This command also has a query form.
Syntax :PAGE:MEASure:QSCV:VAR1:COMPliance compliance | MINimum | MAXimum
:PAGE:MEASure:QSCV:VAR1:COMPliance?
Parameter
Query response compliance <newline><^END>
compliance is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:COMP?"
ENTER @Hp4155;A
Parameter Type Explanation
compliance numeric VAR1 compliance value. –1 to 1 A. The range of
this value depends on the type of VAR1 unit being
used.