5-200 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:QSCV:UNIT
:PAGE:MEASure:QSCV:UNIT
This command specifies the measurement channel for Quasi-static CV
measurements. This command also has a query form.
At *RST, this value is set to DEF.
Syntax :PAGE:MEASure:QSCV:UNIT SMU1 | SMU2 | SMU3 | SMU4 | SMU5 | SMU6 |
DEFault
:PAGE:MEASure:QSCV:UNIT?
Parameter
Query response SMU1 | SMU2 | SMU3 | SMU4 | SMU5 | SMU6 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:UNIT DEF"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:UNIT?"
ENTER @Hp4155;A$
See also :PAGE:CHANnels:CDEFinition:SMU:FUNCtion
:PAGE:CHANnels:CDEFinition:SMU:MODE
Parameter Type Explanation
SMUn character SMUn is assigned as the measurement channel,
where n is integer (1 to 6).
DEFault character VAR1 channel is assigned as the measurement
channel.