5-222 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:DELay
:PAGE:MEASure[:SWEep]:DELay
This command sets the DELAY TIME of SMU.
At *RST, this value is 0.
Syntax :PAGE:MEASure [ :SWEep] :DELay delay_time | MINimum | MAXimum
Parameter
Query response delay_time <newline><^END>
delay_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:DEL 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:DEL MIN"
OUTPUT @Hp4155;":PAGE:MEAS:DEL?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:DEL? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:HTIMe
This command sets the HOLD TIME of sweep measurement.
At *RST, this value is 0.
Syntax :PAGE:MEASure [ :SWEep] :HTIMe hold_time | MINimum | MAXimum
Parameter
Query response hold_time <newline><^END>
hold_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:HTIM 1.5"
Parameter Type Explanation
delay_time numeric 0 to 65.535 (s). Resolution: 100 ms.
Parameter Type Explanation
hold_time numeric 0 to 655.35 (s). Resolution: 10 ms.

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals