5-92 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:SMU<n>:SRESistance
:PAGE:CHANnels[:CDEFinition]:SMU<n>:SRESistan
ce
This command sets the SERIES RESISTANCE of SMU<n>.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
At *RST, this value is 0 OHM.
Syntax :PAGE:CHANnels [ :CDEFinition] :SMU<n>:SRESistance resistance | MINimum |
MAXimum
Parameter
You can use the suffix of OHM, KOHM, and MOHM.
Query response resistance <newline><^END>
resistance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES 1E4"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES 10KOHM"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES 1MOHM"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES MAX"
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:SRES? MAX"
ENTER @Hp4155;A
:PAGE:CHANnels[:CDEFinition]:SMU<n>:STANdby
This command sets the standby (STBY) function of SMU<n>.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
At *RST, standby function is OFF.
Syntax :PAGE:CHANnels [ :CDEFinition] :SMU<n>:STANdby OFF | ON | 0 | 1
Parameter Type Explanation
resistance numeric 0, 1E4, 1E5, or 1E6

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals