5-192 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:QSCV:CONStant:VSU<n>[:SOURce]
:PAGE:MEASure:QSCV:CONStant:VSU<n>[:SOURce]
This command sets the SOURCE value of the constant source VSU<n> for the
Quasi-static CV measurement. This command also has a query form.
Where <n> is required to specify VSU number. Valid VSU numbers are VSU1 and
VSU2.
Use this command only if the function of the specified VSU is CONStant.
At *RST, this value is set to 0.
Syntax :PAGE:MEASure:QSCV:CONStant:VSU<n>[ :SOURce] source | MINimum |
MAXimum
:PAGE:MEASure:QSCV:CONStant:VSU<n>[ :SOURce]?
Parameter
Query response source <newline><^END>
source is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:VSU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:VSU1?"
ENTER @Hp4155;A
Parameter Type Explanation
source numeric 20 to 20 V