5-226 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR1:COMPliance
:PAGE:MEASure[:SWEep]:VAR1:COMPliance
This command sets the COMPLIANCE value of VAR1.
If the unit type of VAR1 is VSU, this parameter is ignored.
At *RST, this value is 100 mA.
Syntax :PAGE:MEASure [ :SWEep] :VAR1:COMPliance compliance | MINimum |
MAXimum
Parameter
Query response compliance <newline><^END>
compliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR1:MODE
This command sets the SWEEP MODE of VAR1 for normal (not knob) sweep
measurement. At *RST, this value is SINGle.
Syntax :PAGE:MEASure [ :SWEep] :VAR1:MODE SINGle | DOUBle
Parameter
Parameter Type Explanation
compliance numeric -200 V to 200V or -1 A to 1 A. The range of this
value depends on the unit type of VAR1.
Parameter Type Explanation
SINGle character single stair sweep
DOUBle character double stair sweep