5-226 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR1:COMPliance
:PAGE:MEASure[:SWEep]:VAR1:COMPliance
This command sets the COMPLIANCE value of VAR1.
If the unit type of VAR1 is VSU, this parameter is ignored.
At *RST, this value is 100 mA.
Syntax :PAGE:MEASure [ :SWEep] :VAR1:COMPliance compliance | MINimum |
MAXimum
Parameter
Query response compliance <newline><^END>
compliance is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR1:MODE
This command sets the SWEEP MODE of VAR1 for normal (not knob) sweep
measurement. At *RST, this value is SINGle.
Syntax :PAGE:MEASure [ :SWEep] :VAR1:MODE SINGle | DOUBle
Parameter
Parameter Type Explanation
compliance numeric -200 V to 200V or -1 A to 1 A. The range of this
value depends on the unit type of VAR1.
Parameter Type Explanation
SINGle character single stair sweep
DOUBle character double stair sweep

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals