EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #276 background imageLoading...
Page #276 background image
5-194 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:QSCV:HTIMe
:PAGE:MEASure:QSCV:HTIMe
This command sets the HOLD TIME for Quasi-static CV measurements. The hold
time is the time from the start of the first sweep step to the start of the delay time.
This command also has a query form.
At *RST, this value is set to 0.
Syntax :PAGE:MEASure:QSCV:HTIMe hold_time
:PAGE:MEASure:QSCV:HTIMe?
Parameter
Query response hold_time <newline><^END>
hold_time is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:HTIM 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:HTIM?"
ENTER @Hp4155;A
Parameter Type Explanation
hold_time numeric 0 to 655.35 (s). Resolution: 10 ms.

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals