5-104 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:CHANnels:UFUNction[:MENU]
:PAGE:CHANnels:UFUNction[:MENU]
This command changes the present display page to CHANNELS: USER
FUNCTION DEFINITION page.
This command does not have query form.
Syntax :PAGE:CHANnels:UFUNction [ :MENU]
Example OUTPUT @Hp4155;":PAGE:CHAN:UFUN"
:PAGE:CHANnels:UVARiable:CATalog?
This query command returns the number and list of currently defined user variables.
This command has query form only.
Syntax :PAGE:CHANnels:UVARiable:CATalog?
Query response num_variable,variable_name_list <newline><^END>
num_variable is NR1 response data.
variable_name_list is a list of the user variable names separated by commas. The
data type is string response data, but does not contain double quote characters at
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:CHAN:UVAR:CAT?"
ENTER @Hp4155;A,B$