5-96 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:VMU<n>:VNAMe
:PAGE:CHANnels[:CDEFinition]:VMU<n>:VNAMe
This command sets the VNAME of VMU<n>.
<n> is required to specify VMU number. Valid VMU numbers are VMU1 through
VMU2.
At *RST, this value is VMU<n>.
Syntax :PAGE:CHANnels [ :CDEFinition] :VMU<n>:VNAMe name
Parameter
Query response name <newline><^END>
name is the string response data, but does not contain double quote characters at the
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:CHAN:VMU1:VNAME 'VD'"
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:VNAME?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition]:VSU<n>:DISable
This command deletes the settings of VSU<n>.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
This command does not have query form.
Syntax :PAGE:CHANnels [ :CDEFinition] :VSU<n>:DISable
Example OUTPUT @Hp4155;":PAGE:CHAN:VSU1:DIS"
Parameter Type Explanation
name string String of up to 6 alphanumeric characters.
1 st character must be alphabet.

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals