Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-127
SCPI Commands
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:NAME
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:NAME
This command selects the variable name of X, Y1, or Y2 axis.
At *RST, this value is:
Syntax :PAGE:DISPlay [ :SETup] :GRAPhics:X|:Y1|:Y2:NAME var_name
Parameter
Query response var_name <newline><^END>
var_name is string response data, but does not contain double quote characters at
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:DISP:GRAP:X:NAME 'VD'"
OUTPUT @Hp4155;":PAGE:DISP:GRAP:X:NAME?"
ENTER @Hp4155;A$
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:SCALe
This command selects whether the scale of X, Y1, or Y2 axis is LINear or
LOGarithmic.
At *RST, this value is LINear for X and Y1, and not defined for Y2.
Syntax :PAGE:DISPlay [ :SETup] :GRAPhics:X|:Y1|:Y2:SCALe LINear | LOGarithmic
AXIS variable name
XV3
Y1 I3
Y2 not defined
Parameter Type Explanation
var_name string data variable name