EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #209 background imageLoading...
Page #209 background image
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-127
SCPI Commands
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:NAME
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:NAME
This command selects the variable name of X, Y1, or Y2 axis.
At *RST, this value is:
Syntax :PAGE:DISPlay [ :SETup] :GRAPhics:X|:Y1|:Y2:NAME var_name
Parameter
Query response var_name <newline><^END>
var_name is string response data, but does not contain double quote characters at
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:DISP:GRAP:X:NAME 'VD'"
OUTPUT @Hp4155;":PAGE:DISP:GRAP:X:NAME?"
ENTER @Hp4155;A$
:PAGE:DISPlay[:SETup]:GRAPhics:X|:Y1|:Y2:SCALe
This command selects whether the scale of X, Y1, or Y2 axis is LINear or
LOGarithmic.
At *RST, this value is LINear for X and Y1, and not defined for Y2.
Syntax :PAGE:DISPlay [ :SETup] :GRAPhics:X|:Y1|:Y2:SCALe LINear | LOGarithmic
AXIS variable name
XV3
Y1 I3
Y2 not defined
Parameter Type Explanation
var_name string data variable name

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals