3-4 Agilent 4155C/4156C SCPI Command Reference, Edition 1
Status Reporting Structure
Status Byte Register
Table 3-1 Status Byte Register
Bit Definition Explanation
0 Emergency Status Bit
Summary Message
Indicates whether one or more of the enabled
Emergency Status Register bits is set.
1 Measurement/Stress
Status Bit Summary
Message
Indicates whether one or more of the enabled
Measurement/Stress Status Register bits is
set.
2 NOT USED Always zero.
3 Questionable Status Bit
Summary Message
Indicates whether one or more of the enabled
Questionable Status Register bits is set.
4 Message Available
(MAV) Queue Summary
Message
Indicates whether the Output Queue is
empty.
5 Standard Event Status
Bit (ESB) Summary
Message
Indicates whether one or more of the enabled
Standard Events Status Register bits is set.
6 Request Service (RQS)
Message
Indicates whether a SRQ (Service Request)
has occurred.
Master Summary Status
(MSS) Summary
Message
Indicates that the instrument has at least one
reason for requesting service.
7 Operation Status Bit
Summary Message
Indicates whether one or more of the enabled
Operation Status Register bits is set.

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals