EasyManua.ls Logo

Agilent Technologies 8753ET - Page 37

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 2-11
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
6. When all measurements are complete, replace the verification device with the
verification mismatch, as shown in Figure 2-6. Be sure that you connect Port A of the
verification mismatch to reference test port 1.
Figure 2-6 Mismatch Device Verification Setup 1
7. Press .
8. When the analyzer finishes all the measurements, connect the mismatch verification
device as shown in Figure 2-7. Notice that Port B is now connected to reference test
port 1.
Figure 2-7 Mismatch Device Verification Setup 2
9. Press .
10.You have completed the system verification procedure when the analyzer displays
Ver Def 4 DONE.
RETURN
TESTS
30 x1
EXECUTE TEST
CONTINUE
RETURN
TESTS
31 x1
EXECUTE TEST
CONTINUE

Table of Contents

Other manuals for Agilent Technologies 8753ET

Related product manuals