Chapter 2 2-11
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
6. When all measurements are complete, replace the verification device with the
verification mismatch, as shown in Figure 2-6. Be sure that you connect Port A of the
verification mismatch to reference test port 1.
Figure 2-6 Mismatch Device Verification Setup 1
7. Press .
8. When the analyzer finishes all the measurements, connect the mismatch verification
device as shown in Figure 2-7. Notice that Port B is now connected to reference test
port 1.
Figure 2-7 Mismatch Device Verification Setup 2
9. Press .
10.You have completed the system verification procedure when the analyzer displays
Ver Def 4 DONE.