EasyManua.ls Logo

Agilent Technologies 8753ET - Page 69

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 2-43
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
40.Press .
41.Press to put marker 1 at the
minimum magnitude location of the trace.
42.Press to position marker 2 at
the maximum magnitude location of the trace.
43.Write the marker 1 or marker 2 reading, whichever has the largest absolute magnitude,
in the performance test record.
Power Meter Calibration for Test Port 1 from 3 GHz to 6 GHz (Option 006 only)
44.Connect the equipment as shown in Figure 2-24.
Figure 2-24 Setup for Power Meter Calibration on Test Port 1
45.Press
.
46.Press to start the power meter
calibration for test port 1.
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz (Option 006 only)
47.When the analyzer displays the message POWER METER CALIBRATION SWEEP DONE,
connect the equipment as shown as in Figure 2-25.
Meas
INPUT PORTS
A
Sweep Setup
TRIGGER MENU
SINGLE
Marker
MARKER 1
Marker Search
SEARCH:MIN
Marker
MARKER 2
Marker Search
SEARCH:MAX
Sweep Setup
TRIGGER MENU
CONTINUOUS
Meas
INPUT PORTS
TESTPORT 1
Cal
PWRMTR
ONE SWEEP
TAKE CAL SWEEP

Table of Contents

Other manuals for Agilent Technologies 8753ET

Related product manuals