EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
2-54 Chapter 2
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
49.Connect the SHORT end of the open/short combination to the analyzer test port 2.
50.Press .
51.Connect the termination to the analyzer test port 2.
52.Press .
53.When the analyzer displays PRESS ’DONE’ IF FINISHED WITH STD(s), press
.
Wait for the message COMPUTING CAL COEFFICIENTS to disappear from the analyzer
display before proceeding to the next step.
54.Connect the equipment as shown in Figure 2-32.
Figure 2-32 Transmission Calibration Test Setup
55.Press .
56.Press .
Load Match (Forward) Calibration Coefficient
57.Press .
58.When the test is done, press .
59.Using the front panel knob, locate the maximum value of the data trace for the 300 kHz
to 1.3 GHz frequency range.
60.Write the maximum value on the performance test record.
61.Repeat the previous three steps for the other frequency ranges listed on the
performance test record.
Directivity (Reverse) Calibration Coefficient
62.Press .
63.When the analyzer finishes the test, repeat steps 58 through 61.
REVERSE: SHORT
REVERSE: LOAD
STANDARDS DONE
TRANSMISSION
DO BOTH FWD + REV
DONE 2-PORT CAL
System
SERVICE MENU
TESTS
36 x1
EXECUTE TEST
Marker
System
SERVICE MENU
TESTS
38 x1
EXECUTE TEST

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals