C8000 Battery Testing System User Manual
3 Cadex BatteryLab 187
Required Action Sub-
parameters
Description of the Action
calculation using a 1Khz AC
signwave.
DC Impedance:
Performs a resistance
calculation using DC current
as per IEC standards.
I/O ports must be enabled
Similar to REST but allows the reconfiguring of the
conditions of the Auxiliary ports.
Will perform a charge to the battery for Duration of 3
minutes, in an attempt to raise the battery voltage.
This function will disable many of the
safe guards. Use with EXTERME
CAUTION.
All functions are suspended until the user prompts to
continue the service by selecting “RESUME”
Dchg Device
Selection of the Device to
perform a discharge
Uses a playback waveform to perform a discharge on the
C8000 or on a supported external load.
Requires a supported Electronic DC Load
See section 2.12.1.1 Programmable DC Electronic
Loads
Start:
Indicates the starting
temperature for the profile.
End:
Indicates the target
temperature to be reached.
Interval:
Controls the temperature rise
of the chamber.
To calculate the temperature
raise per minute.
(End – Start)/interval
(minutes)
Deviation:
The error allowed to the
target temperature.
Type:
Selects the support thermal
Requirement:
Requires a supported Thermal Chamber
See section 2.12.1.2 Thermal Chambers
Controls the temperature environment of a supported
thermal Chamber.