C8000 Battery Testing System User Manual
Safety and Warranty 9
1. From the Tools menu, click System Update.
2. Select the check box for each C8000 to restore calibration
data for.
3. Under the Calibration tab, click Restore Calibration Data
to restore previously backed up data.
4. Click OK.
Important
Each new C8000 has been calibrated at the factory,
and has its factory calibration data stored on the
C8000. This calibration data is essential to providing
accurate testing of batteries.
On receipt of each new C8000, configure the unit with
a static IP address, the network mask address, and the
PC Host IP address. Then run BatteryLab and
configure BatteryLab to allow connection from the new
C8000. Choose Connect to allow BatteryLab to
automatically backup the calibration data for the new
C8000.
This calibration data is essential to providing accurate
testing of batteries. After installing and configuring
Cadex BatteryLab to communicate with the C8000, use
BatteryLab to back up the calibration data to the
computer.
Each new C8000 should be connected to the PC for
BatteryLab to automatically back up the calibration
data.
If, for any reason, the calibration data on the C8000
gets corrupted, the calibration data can be restored
from the backup copy. Otherwise, the C8000 must be
shipped to a Cadex Service Centre for re-calibration.
The shipping and service cost is not covered by the
warranty.