C8000 Battery Testing System User Manual
280 5 Status Messages
A battery has been inserted and is ready to
be serviced. (Also, indicated as Startup).
Battery is being reconditioned.
The program will proceed to the next
step when recondition has completed.
Channel is in a rest period, as specified in
the program.
The program will proceed to the next
step when resting period is complete.
An abort command has been issued. Program
terminated.
The user issued an abort command.
SYSTEM TEMP.
HIGH: COOLING
Service on all channels is temporarily
suspended due to high temperature inside
the tester.
Service will resume in a few minutes,
after the board has cooled down. If it
reoccurs continuously, move the tester
to a cooler room.
TARGET
CAPACITY NOT
MET after service
Final capacity of the battery is below the
target capacity. Attempts to improve battery
capacity did not succeed in bringing capacity
above target.
Battery is old and has less operating
time than manufacturer’s specification.
The program is running a trickle
charge.
The reconditioning process is complete
and the battery is being recharged.
Battery voltage is too high. Processing
has ended.
Battery may be a high capacity
type battery.
Check that correct contacts are
used.
Battery may be in poor condition.
The time programmed in a custom
program setup has completed.
The program goes to the next step.