EasyManua.ls Logo

Cadex C8000 - Page 39

Cadex C8000
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C8000 Battery Testing System User Manual
Battery Tester 27
Battery Compatibilty
Chemistries
All Standard Battery Types
Lead Acid
Ni-based: NiMH, NiCd
Li-based: Li-ion, Li-Phosphate
Capacity Range
50mAh to 1000Ah
Electrical
Input Power
100-240 VAC
50-60 Hz
Battery Voltage Range
Nominal 1.2V-36V (max 45V) supports serial
connection of:
Up to 10 Li-ion cells (3.6V/cell)
Up to 18 Lead Acid cells (2.0V/cell)
Up to 30 Nickel-based cells (1.2V/cell)
Maximum Charge Power
100W per channel = 400W total
Maximum Discharge Power
80W per channel = 320W total
Dual Power Port cable may be to increase the
maximum Power Rate to 160W per channel pair
External Load may be used to increase
Discharge Rate up to 2400W
Maximum Charge/Discharge
Current
30mA to 10A set in 1mA increments
Dual Power Port cable may be to increase the
maximum Discharge Rate to 20A per channel
pair
External Load may be used to increase
Discharge Rate up to 240A
Output Frequency
Square 2 kHz (500 microseconds)
C8000 Accuracy
Voltage = 0.1% full scale
Current = 0.25% full scale
Other Specifications
Power Management
Current automatically scaled down if power
limits are exceeded message indicates scaled
test
Charge Methods
Charge methods are dependent upon chemistry:
Li-ion + Lead Acid = constant current/

Table of Contents