EasyManua.ls Logo

Edwards HemoSphere

Edwards HemoSphere
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
210
HemoSphere Advanced Monitor E Computation Constants
* To optimize cardiac measurement, it is recommended that the temperature of the injectate correspond to
one of the temperature ranges listed in the catheter’s directions for use.
Table E-2 Computation constants for in-line temperature probe
Injectate
temperature
range* (°C)
Injectate
volume
(mL)
Catheter size (French)
8 7.5 7 6 5.5
Room temp.
22.5–27 °C
10
5
0.601
0.294
0.599
0.301
0.616
0.311
0.616
0.307
0.624
0.310
Room temp.
18–22.5 °C
10
5
0.593
0.288
0.593
0.297
0.603
0.295
0.602
0.298
0.612
0.304
Cold (iced)
5–18 °C
10
5
0.578
0.272
0.578
0.286
0.570
0.257
0.568
0.276
0.581
0.288
Cold (iced)
0–5 °C
10
5
0.562
0.267
0.563
0.276
0.537
0.217
0.533
0.253
0.549
0.272
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14