210
HemoSphere Advanced Monitor E Computation Constants
* To optimize cardiac measurement, it is recommended that the temperature of the injectate correspond to
one of the temperature ranges listed in the catheter’s directions for use.
Table E-2 Computation constants for in-line temperature probe
Injectate
temperature
range* (°C)
Injectate
volume
(mL)
Catheter size (French)
8 7.5 7 6 5.5
Room temp.
22.5–27 °C
10
5
0.601
0.294
0.599
0.301
0.616
0.311
0.616
0.307
0.624
0.310
Room temp.
18–22.5 °C
10
5
0.593
0.288
0.593
0.297
0.603
0.295
0.602
0.298
0.612
0.304
Cold (iced)
5–18 °C
10
5
0.578
0.272
0.578
0.286
0.570
0.257
0.568
0.276
0.581
0.288
Cold (iced)
0–5 °C
10
5
0.562
0.267
0.563
0.276
0.537
0.217
0.533
0.253
0.549
0.272
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14