6.2. Spectrum FLIR Detection
The current spectrum is not cleared before recording data, that is, new measurements
are added to the existing spectrum. To record a pristine spectrum, use the
Clear com-
mand irst (see below).
The acquisition of spectrum data stops in one of the following events:
• You end the acquisition with
Stop .
• The end of the real or life duration you speciied (see 6.8, p. 121) is reached. You know
the duration is preset because of the indicator (Figure 111, p. 108) and the timer is
counting down.
emaining Acquisition Duration
Figure 111. Acquiring spectrum data while the duration is preset
• You switch the FLIR identiFINDER R300 to Identiication Mode (see 4.3, p. 90). This dis-
cards the current spectrum and loads the spectrum from the identiication process into
the spectrum viewer.
• You
Exit the spectrum view.
Exit
Return to the superior menu.
If you try to exit during data acquisition, you will be asked to conirm the command
because the acquisition will stop.
Spectrum Cursor
You can place a cursor for detailed inspection along the spectrum’s horizontal axis. The cur-
sor position is represented on screen by a vertical line with a down-pointing triangle on top
(Figure 112, p. 109).
The number of counts at the cursor’s position is shown above the triangle together with the
position as channel number (Figure 112, p. 109) or energy (Figure 113, p. 110).