FLIR Detection 4. FLIR identiFINDER R300 Modes
Settings and Commands
..
Switch Off
Switch off automated mode.
Depending on your variant, this might require entering the password (2.2.7, p. 46).
..
Cancel
Cancel this measurement. The next measurement will start as scheduled.
Detected Nuclides
Once the Automated Mode measurement is complete, the FLIR identiFINDER R300 will re-
port any nuclide(s) identiied and indicate the reference number assigned to the stored
results.
..
30 s
Initially, the only available command on the nuclide results screen shows a countdown timer
starting at 30 s. When the timer expires, the FLIR identiFINDER R300 will return to the
screen that was active before the Automated Mode measurement started. Pressing
..
30 s
prior to the timer expiring will cancel the timer and the command will change to
..
Exit .
..
Exit
Only appears if the results countdown timer was cancelled. Press to close the results screen
and return to the screen that was active before the Automated Mode measurement started.
Results Storage
The results of the measurements are saved in the database of the FLIR identiFINDER R300.
They are marked as being results of automated measurements (see 8.11, p. 199).
4.5 Remote-Controlled Spectrum Acquisition
While the FLIR identiFINDER R300 is connected to a computer, you can collect and analyze spectra
via the web interface. This method of data acquisition for spectra interrupts other operations of
the instrument.
Alarms will still be reported, if necessary.
During the acquisition, a message is shown (Figure 99, p. 96). You can stop it on the instrument
with
..
Stop or via the web interface with
.
.
.
Stop (see 8.16, p. 209).
After the acquisition, the FLIR identiFINDER R300 resumes the mode it was in before. If the previ-
ous operation was interrupted during a time limited task like collecting background or identiica-
tion data, it must start over again.