FLIR Detection 4. FLIR identiFINDER R300 Modes
Identic
ation Condence
ation Condence
ation Condence
Nuclide
Names
Names
Names
Nuclide
Usage
Nuclide
Severity
R
eference Number of
the Saved Record
Figure 96. Identiication Mode: Identiied radionuclides
Continue
Acquire more data to improve the identiication. The acquisition continues for the duration
you speciied (see 6.4, p. 116) no matter whether you limited the duration or the minimum
number of counts for the initial acquisition.
Spectrum
Inspect the acquired spectrum.
This command is available only after you entered the password or if the password pro-
tection is off.
Please refer to 6.2, p. 106 for options and commands for spectrum viewing. After you return
from the spectrum, you can
Continue to acquire data, or Exit the identiication mode.
4.4 Automated Mode
In automated mode, the FLIR identiFINDER R300 collects data for a radionuclide identiication at
speciied time intervals or in response to a dose rate warning. You can switch this mode on or off
and specify the interval (see 6.20, p. 135).
For each measurement, data will be collected until an instrument-speciic number of gamma counts
has been registered (cf. “Dynamic” identiication duration, see 6.4, p. 116).
In low-radiation environments, this can take quite a while, even longer than the interval you
speciied. The next scheduled measurement will be omitted in such cases.
The start of an automated measurement will be announced 60 s in advance by a countdown shown
instead of the current time on the display of the FLIR identiFINDER R300 (Figure 97, p. 94).
identiFINDER
®
R300/en/2014.4(13623)/Feb2015 93