FLIR Detection 6. FLIR identiFINDER R300 Menu Reference
This screen (Figure 123, p. 117) groups options that inluence the identiication of nuclides (see
4.3, p. 90).
Figure 123. Upper items of the identiication settings
Settings and Commands
Duration
Specify the duration of data acquisition for the nuclide identiication. Allot more time to the
identiication measurement if:
• you want to identify a weak source.
• you are at a place with high background radiation.
• you want to identify a mixture of radionuclides.
• you want to identify a shielded source.
You can specify the durations “Dynamic” or from 1 s to 999 s (16 m 39 s).
If you specify a duration of “Dynamic”, the FLIR identiFINDER R300 will collect data un-
til 4800 gamma counts have been registered. This may take a while especially if the
radiation level is near the background level.
Extra Duration
Specify the duration of continued data acquisition for the nuclide identiication. This dura-
tion is applied when you use the
Continue command in identiication mode (see 4.3, p. 90),
no matter whether you speciied the duration or the minimum number of counts above.
You can specify an extra duration from 1 s to 999 s (16 m 39 s).
Condence
Specify if you want the level of conidence of a nuclide identiication shown or hidden in iden-
tiication results. The conidence is expressed on a scale from 0 (poor) to 10 (excellent).
identiFINDER
®
R300/en/2014.4(13623)/Feb2015 117