A.1. identiFINDER R 300: Specications FLIR Detection
continuation
Dose Rate Range 100 nSv/h 10 mSv/h (10 μrem/h
1.0 rem/h);
Dose Rate Accuracy (
137
Cs) ±30 %
Overload Threshold 5 mSv/h
Neutron Sensitivity
..
B 2.6 cps/nv; ±20 %
Nuclide Identiication According to ANSI N42.48
Typical Resolution 3.5 % FWHM at 662 keV at 25 °C (77 °F)
ambient temperature
Service
Recommended Interval 5 a
Physical
Dimensions (W × D × H) 70.3 mm (2.77 in) × 33.8 mm (1.33 in) ×
125.5 mm (4.94 in)
Weight
..
A 340 g (11.99 oz)
Weight
..
B 370 g (13.05 oz)
Housing Material Aluminium
Environmental
Operating Temperature -20 °C +50 °C (-4 °F +122 °F)
Storage Temperature +10 °C +35 °C (+50 °F +95 °F)
Temperature Change Temperature change must not exceed 8.4 °C
(15.1 °F)/min
Relative Humidity 10 % 93 %, non condensing
Shock Resistance According to ANSI N42.48-2008
Protection Rating IP63 according to IEC 60529
Battery
Type Internal (Li-ion, single cell, rechargeable)
Capacity 1950 mAh; 3.7 V; 25 °C (77 °F)
Battery Life 300 500 full charge-discharge cycles; 3 a 5 a
Charging Duration 3 h at 25 °C (77 °F) when instrument off and
connected to AC power charger and >5.5 h when
powered via USB.
Operating Duration 24 h at 25 °C (77 °F) in dose rate mode with
dimmed display back light and GPS switched off
Display
Type Translective color LCD
to be continued…
238 identiFINDER
®