Chapter 10: Set Preferences
Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) 10-7
Operator Manual © 2008 General Electric Company. All rights reserved.
Perform C-Echo Test
Use this function when you want to check to see if the system is communicating with a particular
network host.
1. Select a host from the Network Hosts list.
2. Press [C-ECHO TEST].
A message appears to notify you if the test passed or failed.
A “passed” message means that the network host is working and that you can retrieve exams
from it or transfer exams to it.
Failed C-Echo Test
A “failed” message means that the system could not contact the network host.
Perform the following tasks to resolve the problem:
Re-try the test at a later time. The host may be temporarily unavailable.
Confirm the host configuration on the Edit screen.
If the problem persists, contact your technical support group or system administrator.
Printers
Follow this process to access the Network Connections screen.
1. On the Worklist screen, press [UTILITIES].
The System-System screen appears.
2. Press [NETWORK CONNECTIONS].
The Network Connections screen (Figure 10-5) appears.
Figure 10-5 Network Connections screen
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Definium 5000 and is the answer not in the manual?

GE Definium 5000 Specifications

General IconGeneral
BrandGE
ModelDefinium 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals