EasyManua.ls Logo

GE Definium 5000 - Correct Pasting

GE Definium 5000
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix A: Image Pasting
Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) A-10
Operator Manual © 2008 General Electric Company. All rights reserved.
The registration markers (brackets) show the general area of pasted overlap (Figure A-17). Use the Image
Magnifying Glass function to review registration quality.
Figure A-17 Close-up of registration marker and mis-registered area
Follow this process to paste images.
1. From the tool selection list, select Reprocessing.
2. Press the [PASTE] button.
NOTE: It may take a few moments for images to appear.
3. Review the pasted image.
If the image is satisfactory, press [APPLY TO IMAGE] to save the composite and return to the main
Image Viewer screen.
If the pasted image is not satisfactory, follow the steps described in Correct Pasting
(p. A-10) to
make adjustments.
NOTE: The pasted composite will not be saved until [APPLY TO IMAGE] is pressed.
Correct Pasting
Follow this process to adjust a pasted image.
1. Select a region to re-paste from the drop-down list (Figure A-18). For example: 1/2, 2/2.
[RESET REGION] removes any adjustment from the selected region.
Figure A-18 Select region drop-down list.
2. Press and drag a red cross-hair to move it (Figure A-19).
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).