COMMISSIONING
94 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Name Value Range Default Value Explanation
[SAT frequency limit] <120>
Determines the
maximum frequency
change when the
displacer has been
raised.
[SAT intended interface] <Interface 1>
<Interface 2>
<Interface 3>
<Interface 1>
Determines at/around
which interface the SAT
is performed.
[SAT invalidate level] <enable>
<disable>
<disable>
Determines if the actual
interface will go the
health BAD during the
SAT.
[SAT level last fail] 999.9999
Shows the level at
which the last SAT
failed.
[SAT level last executed]
Shows the interface
level at the moment the
SAT was performed.
[SAT minimum level
executed]
Shows the minimum
interface level where
the SAT was
performed.
[SAT maximum level
executed]
Shows the maximum
interface level where
the SAT was
performed.
[SAT highest number of
retries]
Shows the number of
retries in case a SAT
performance failed.
[SAT number of
executed tests]
Shows the number of
SAT’s performed since
the gauge was
powered up.
[SAT status]
Shows the result of the
last performed SAT.
[SAT time elapsed since
last SAT]
Shows the elapsed
time since the previous
SAT
[SAT RTC last fail]
Shows the date of the
RTC when there was a
SAT fail.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals