COMMISSIONING 
 
 
234  Service Manual    Part No.: 4417340 Revision 1 
  954 SmartServo FlexLine
 
  Level calibration using the top of ball valve 
 If level dipping is not possible, the 954 SmartServo FlexLine can be 
calibrated using the top of a ball valve as reference point.  
 
 
 
Make sure the servo auto test downwards is disabled, else the servo gauge will end 
up in a SAT failure and will stop measuring level. Refer to section 8.7 
Proceed as follows: 
Entity name  Entity name 
(SmartView) 
Description   
[Calibrate] [Calibrate] 
(Product level) 
The displacer will be raised until it stops against the 
flange of the level gauge. 
Caution  
If a 45 mm displacer is used, immediately stop the 
Calibrate command as soon as the displacer is above 
the ball valve, followed by a freeze command. 
Make sure the displacer is positioned above the ball 
valve. Close the ball valve. 
[Unlock] [Unlock] 
(Product level) 
Unlock the gauge and wait the displacer reaches the top 
of the ball valve.   
Calculate the immersion depth of the displacer at the 
product interface (for a 90 mm displacer use 3 mm; for a 
45 mm displacer use 12 mm). 
[Reference level]  [Reference level] 
(Product level) 
Enter the position of the top of the ball valve with respect 
to the tank zero adding the immersion depth of the 
displacer at the product level. 
[Accept reference]  [Accept reference] 
(Product level) 
By giving this command, the level value entered in entity 
[Reference level] is accepted as level value and will be 
shown on the display after the automatic reset. 
[Calibrate] [Calibrate] 
(Product level) 
The displacer will now raise from the ball valve. Let it 
stop against the flange or give a freeze command.   
Open the ball valve. 
[Unlock] 
[Unlock] (Product 
level) 
Unlock the gauge. The displacer will now move down till 
it reaches the level 
 
For verification of the level calibration, refer to section 8.8.2