EasyManua.ls Logo

Honeywell FlexLine SmartServo 954 - Operational Commands

Honeywell FlexLine SmartServo 954
304 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMISSIONING
Part No.: 4417340 Revision 1 Service Manual 75
954 SmartServo FlexLine
7.3.4.9 Operational commands
Name Value Range Default Value Explanation
[Host command mode] <Enable>
<Disable>
All <Enable>
With this entity the user
can enable or disable
the following
commands:
-Interface 1
-Interface 2
-Interface 3
-Block
-Calibrate
-Dip mode
-Freeze
-Interface profile
-Lock test
-Test gauge
-Tank profile
-Unlock
-Density water dip
-Water density dip
-Test
-Lock to level
-Lock test to level
[Freeze]
This command stops
the motor. This
situation will be
maintained until a
[Unlock] command is
given.
[Test gauge]
This command will rise
the displacer 6 cm. up
and after that the
interface is searched
again. (reproduction
test)

Table of Contents