OPERATION 
 
 
254  Service Manual    Part No.: 4417340 Revision 1 
  954 SmartServo FlexLine
 
  Verify level calibration on top of ball valve   
If the top of the ball valve is used as reference point, the level calibration 
of the 954 SmartServo FlexLine can be verified.  
 
 
Make sure the servo auto test downwards is disabled, else the servo gauge will end 
up in a SAT failure and will stop measuring level. Refer to section Error! Reference 
source not found.. 
 
 Proceed as follows: 
Entity name  Entity name 
(SmartView) 
Description   
[Calibrate]  
[Calibrate] 
(Product level) 
The displacer will be raised until it stops against the flange 
of the level gauge. 
Caution  
If a 45 mm displacer is used, immediately stop the 
Calibrate command as soon as the displacer is above the 
ball valve, followed by a freeze command. 
Make sure the displacer is positioned above the ball valve. 
Close the ball valve. 
[Unlock] 
[Unlock] 
(Product level) 
Unlock the gauge and wait the displacer reaches the top 
of the ball valve.   
[Reference level] 
[Reference level] 
(Product level) 
Request for the reference level value, which was 
established during calibration. The level reading from the 
gauge should not differ more than ±2 mm (±1/16") with the 
value in entity [Reference level]. 
[Calibrate] 
[Calibrate] 
(Product level) 
The displacer will now raise from the ball valve. Let it stop 
against the flange or give a freeze command.   
Open the ball valve. 
[Unlock] 
[Unlock] 
(Product level) 
Unlock the gauge. The displacer will now move down till it 
reaches the level 
 
If the reference level value and the gauge reading differ more than the 
specified value, the calibration procedure as described in section 7.6.3 
should be repeated.