COMMISSIONING
Part No.: 4417340 Revision 1 Service Manual 143
954 SmartServo FlexLine
7.3.8.4.4 Commissioning Parameters for All Types of Probes
Engauge *Product temperature* Tab
The following entities can be set set by Engauge or HART
SmartView for a specific functioning of the FII-RTD module in an
instrument. The default values are available after initialization of the
non-volatile memory; so check each entity for its correctness.
Name Explanation Value Range Default
[Product immersion depth]
The minimum required
product level distance
above an element before
this element is taken into
account in the average
product temperature
calculation
floating point number:
<-x.x .. +x.x>
<0.5>
[Hysteresis]
This is a hysteresis
mechanism distance
around the switching points
of the elements that are
taken into account in the
calculation.
floating point number:
<-x.x .. +x.x>
<0.1>
[High high alarm]
[High alarm]
[Low alarm]
[Low low alarm]
4 thresholds for activating a
related alarm status in the
Primary Value
floating point numbers:
<-x.x .. +x.x>
<1.0E22>
[Alarm test enable]
Enables (if activated, see
next listed entity) the
simulation of one of the 4
alarms for a minute, by
simulating the actual
measured Primary Value is
slightly below or above the
alarm threshold.
<Enable>
<Disable>
<disable>
[Alarm test]
Activates and selects at the
same time the Alarm test (if
enabled with the [Alarm test
enable] listed before, and the
4 alarm thresholds are
properly set).
<No Alarm>
<High-high alarm>
<High alarm>
<Low alarm>
<Low-low alarm>
<No Alarm>
[Function identification] The current module’s
function name. This
function is visible on the
HART SmartView
d
i
l
13 characters
<Product temp.>
[Element offset]
Each element can
individually be given an
offset.
floating point numbers:
<-x.x .. +x.x>
<0.0>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals