COMMISSIONING
224 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Name Value Range Default Value Description
[Contact 2
hysteresis]
< 0 – 3.402823 × 10
38
>
(meters)
< 0.010 > A value relative to the
threshold that will determine
the point at which
deactivation of a contact 2
safe state occurs after
activation by an overfill or
underfill condition.
For an underfill, deactivation
occurs when the product
level is above (threshold +
hysteresis).
For an overfill, deactivation
occurs when the product
level is below (threshold –
hysteresis).
[Contact 2
safety function
status]
(RO)
Status:
< H >, < O >, < U >, < W >
Status Code:
Defined in Section 7.3.13.1
Status:
< H >
Status Code:
NO ERROR
H: Healthy
O: Overfill condition
U: Underfill condition
W: Warning condition, safe
state is active or proof test is
overdue
[Contact 2
safety shutdown
counter]
(RO)
< 0 – 3.402823 × 10
38
>
< 0 > The number of contact 2 safe
state activations that have
occurred since
commissioning.
[Contact 2
safety shutdown
reason]
(RO)
Safety Function Status Code NO ERROR The safety function status
code identifying the reason
for the last time a contact 2
safe state activation
occurred.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals