MAINTENANCE
272 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Switch off the wire protection before calibrating the force transducer:
Entity name Entity name
(SmartView)
Default
Value
Description
[Wire tension
protection]
[Wire tension pr]
(Product level)
Enable
Disable the wire-protection. This prevents
that under certain circumstances the 25 g
calibration will be aborted.
[Wire rupture
protection]
[Wire rupture pr]
(Product level)
Enable
Disable the wire-rupture. This prevents
that under certain circumstances the 25 g
calibration will be aborted.
[Reset board] [Reset TII-SRV]
(board)
Reset the TII-SRV, which de-activates the
wire rupture.
Remove the displacer from the measuring wire and attach the smallest
test weight of 25 grams.
Entity name Entity name
(SmartView)
Description
[Lock test] [Lock test]
(Product level)
Raise the displacer as high as possible until it is in
full view, and good accessible.
[Freeze] [Freeze]
(Product level)
Send a freeze command to stop the displacer
movement. Correct, if necessary temporarily the
motor limit high setting, in such a way that the gauge
will automatically return to this level after a lock-test
command.
[Balance test]] [Balance test]
(Product level)
Start a balance test. The 954 SmartServo FlexLine
will measure and calculate an average frequency at
25 g over one full drum revolution. Refer to section 0
for more details on performing the balance test.
[Balance test
result average
frequency]
[Avg meas freq [Hz]]
(Product level)
View the average measured frequency after
completion of the balance test and note this value for
[Frequency 0]. The display will show a FR (freeze) in
the lower right corner of the display.
Do not yet reprogram this value in the gauge. Reprogram all values
after completion of all the four calibration measurements. Repeat this
calibration from for the other 3 test weight combinations:
for Frequency 1 with: (25+75 g),
for Frequency 2 with: (25+150 g),
for Frequency 3 with: (25+225 g).