6 - 6
HAPSITE ER Operating Manual
6.2.2  Quick Reference SOP — Running Survey Mode
1 Double-click the ER IQ software icon.
2 Double-click the Run Method icon.
3 Wait for heaters to reach the setpoint temperatures.
4 Click the RUN button in the pop-up window.
5 Sample background for one minute and note the TIC.
6 Hold the probe over the sample until a response that is 2 to 3 times the baseline 
is observed. If the TIC does not increase, sample for a full minute.
7 Press Stop to stop the method.
NOTE: This is a tentative identification. To confirm results, run an Analyze 
(GC/MS) method.
Do not touch the sample with the probe. Do not allow 
liquids to enter the probe.
6.3  ANALYZE (GC/MS) Mode with the Concentrator
6.3.1  Tri- Bed Concentrator
The Tri-Bed concentrator is used for analyzing samples with concentrations in the 
part per million to high part per trillion range. Two default qualitative methods are 
ER_Tri-Bed_PPM_Standard and ER_Tri-Bed_PPB Standard. Use the 
ER_Tri-Bed_PPM_Standard method when a response is seen in Survey. If a 
compound is suspected, but Survey does not show an increase in TIC (total ion 
count, which is a measure of response), use the ER_Tri-Bed_PPB_Standard 
method.
The concentrator feature has increased sensitivity. Use 
the appropriate method to avoid saturating HAPSITE ER.