4 - 1
HAPSITE ER Operating Manual
Chapter 4
Operating HAPSITE ER in Portable Mode
4.1 Starting HAPSITE ER in Portable Mode
Portable Mode refers to using HAPSITE ER without the Laptop computer.
Required Materials
HAPSITE (Analytical Module)
Internal Standard Gas Canister
Carrier Gas Canister
Charged Battery
AC to DC Power Converter Power Supply
Probe
Procedure
1 Assemble HAPSITE ER as shown in section 3.2, Basic Assembly, on page 3-5.
2 Press the POWER button on the front panel. The word POWER will illuminate.
Powering on HAPSITE ER takes one (1) to two (2) minutes. (See Figure 4-1.)
Figure 4-1 Power Switch
NOTE: Power on HAPSITE ER while connected to AC power. Using battery
power to turn on and heat HAPSITE ER will consume over 40% of the
battery’s charge.
3 HAPSITE ER will boot in approximately one minute and will sense which
sample configuration (i.e., concentrator) has been installed. It will begin to
prepare the default method for this sample configuration.