10 - 19
HAPSITE ER Operating Manual
Figure 10-21 Method Editor Inlet Page: GC Temperature Profiles
Adjustments to the Hold Time, Ramp Rate and Temp columns will automatically
update dependent parameters. For example, increasing the Temp will increase the
Ramp Time and increasing the Hold Time will adjust the Start time of the next
parameter. See Figure 10-22.
Figure 10-22 Changing GC Temperature Profiles
NOTE: A maximum of four lines is permitted in this section.
10.6.3 Scan Events
The items displayed in the Scan Events field is dependent upon the type of
method.
Figure 10-23 Scan Events For Probe and SPME
The Filament Delay delays the turning on of the filament. This protects the filament
by allowing the components of the air peak or solvents to pass through the Mass
Spectrometer.