11 - 20
HAPSITE ER Operating Manual
Conc/Factor. . . . . . . . . . . . . . . . Data file concentration of analytes if Global
is selected, or concentration multiplier if
Analyte is selected.
Selection . . . . . . . . . . . . . . . . . . If checked, file will be processed upon
clicking Start.
Display. . . . . . . . . . . . . . . . . . . . . . . Displays the chromatogram for the selected
data file.
Reset Library . . . . . . . . . . . . . . . . . If checked, the calibration curve will be reset.
All points currently contained in the library
will be deleted.
11.3.3 Peak Search
Search . . . . . . . . . . . . . . . . . . . . . . . When Build/Edit is selected, Search
performs a peak detection and integration on
the selected files. When Calibrate is
selected, Search calibrates the library and
calculates the response factors.
Recalculate . . . . . . . . . . . . . . . . . . . Recalculates the peak areas and response
factors without performing a peak search.
This is most useful after manually editing the
baseline points of the peak.
Start . . . . . . . . . . . . . . . . . . . . . . . . . Initiates the Search or Recalculation.
Build/Edit Template . . . . . . . . . . . . Build/Edit must be selected if an analyte
search, deletion of analytes or changing
template parameters is desired.
Calibrate Library. . . . . . . . . . . . . . . Calibrate Library must be selected in order a
calibration curve to be created with the
desired data files.
Search for Analytes . . . . . . . . . . . . Enables a search to be performed on the
selected data file(s) without adding the
detected analytes automatically to the library.
This allows the data to be previewed before
adding it to the library template.
NOTE: When adding compounds to an existing library or Template, use Search
for Analytes. If using Build/Edit Template, the original template will be
overwritten by the new search.