9 - 27
HAPSITE ER Operating Manual
Status . . . . . . . . . . . . . . . . . . . . . . . Status of the mass peak.
OK . . . . . . . . . . . . . . . . . . . . . . . Within minimum and maximum values.
OK LOW. . . . . . . . . . . . . . . . . . . Outside of minimum value but within
acceptable tolerance.
OK HIGH . . . . . . . . . . . . . . . . . . Outside of maximum value but within
acceptable tolerance.
LOW . . . . . . . . . . . . . . . . . . . . . . Below minimum value; needs adjustment.
HIGH . . . . . . . . . . . . . . . . . . . . . Above maximum value; needs adjustment.
FAILED . . . . . . . . . . . . . . . . . . . Cannot located mass peak within window.
9.4.15 Scan Window Menu
Place the mouse cursor in the Peak Scan or Full Scan window and right-click to
display the menu shown in Figure 9-41.
Figure 9-41 Scan Window Options
Undo . . . . . . . . . . . . . . . . . . . . . . . . Returns the screen to its previous state.
Mass Centered Width. . . . . . . . . . . Width in amu that correctly aligns the
calibration peak on the mass axis.
Lock Cursor to Peak . . . . . . . . . . . Locks the cursor to the mass peak to adjust
the mass position.
Manual Scale . . . . . . . . . . . . . . . . . Allows the mass peak windows to be set to a
user defined scale.
Common Scale . . . . . . . . . . . . . . . . Sets the mass peak scan windows to a
common scale based on mass 117.
Show Tune Parameters . . . . . . . . . Displays the EM voltage, Ionizer Control,
Baseline, Threshold and Rod Polarity
settings on the Control Panel.
Show Mass Tune Controls . . . . . . Displays the mass tune controls on the mass
peak scan windows.