6 - 2
HAPSITE ER Operating Manual
6.2 Survey Mode
Survey mode is used for quick analysis and tentative results. It is generally two
minutes long and detects compounds with a concentration greater than 1 ppm.
Do not touch the sample with the probe. Do not allow
liquids to enter the probe.
6.2.1 Sampling Procedure
1 Open the ER IQ software by double-clicking on the ER IQ icon.
(See Figure 6-2.)
Figure 6-2 ER IQ Icon
2 If the probe is attached, HAPSITE will begin preparing a Survey method. If this
method is not the desired one, see section 6.4, Selecting a New Method, on
page 6-11.
NOTE: The following message will be displayed while HAPSITE prepares for
sampling. No action is required from the user. (See Figure 6-3.)
Figure 6-3 NEG Full Power
3 As part of the preparation, an AutoTune will run. If the AutoTune is successful,
the following message will display. Click OK. (See Figure 6-4.) If AutoTune fails,
see section 9.4, Performing Manual Tune, on page 9-7.
Figure 6-4 AutoTune Complete