HAPSITE ER Operating Manual
Clicking the Library Search Parameters button on the Method Editor Search
page will display the following window. See Figure 10-50.
Figure 10-50 Library Search Parameters Window
10.10.3 Peak Search
The peak search section is compromised of the parameters that are used in
distinguishing a peak from the baseline.
Search Window. . . . . . . . . . . . . . . . This value defines the acceptable retention
time range for a peak. The default value is 20
seconds. When using the default value, the
window is 10 seconds on either side of the
expected retention time in order for the
software to make an identification
Min RIC Area . . . . . . . . . . . . . . . . . . The area of the intensity of the largest mass
fragment of the peak must be above this
setpoint.
Min TIC . . . . . . . . . . . . . . . . . . . . . . The area of the intensity of the total ion count
for the peak must be above this setpoint.