10 - 20
HAPSITE ER Operating Manual
If the Filament delay is too short, the high pressure burst
caused by a solvent peak may shut down the HAPSITE
ER and stop the analysis.
The Run Time is the amount of time that the method will run.
The type of concentrator being used is selected in the Concentrator box. The
options are None, Tenax or Carbon. See section 4.4.2, Tenax Concentrator, on
page 4-25 and section 4.4.1, Tri-Bed Concentrator, on page 4-25 for more
information on differences between concentrators.
10.6.4 Headspace Flow Parameter
The Scan Events field for Headspace has an additional parameter, Headspace
Flow Pressure. Headspace Flow Pressure controls the flow rate of carrier gas
through the HSS during the Sample and Purge states. This parameter is only
available when the creating HSS methods.
Figure 10-24 Headspace Flow Parameter
10.6.5 SituProbe Flow Parameter
The Scan Events field for the SituProbe has an additional parameter, SituProbe
Flow Pressure. SituProbe Flow Pressure controls the flow rate of carrier gas
through the SituProbe during the Sample and Purge states. This parameter is
only available when creating SituProbe methods.