9 - 25
HAPSITE ER Operating Manual
9.4.13 Tune and Mass Calibration Status
The Tune & Mass Calibration Status Panel is shown in Figure 9-38.
Figure 9-38 Tune and Mass Calibration Status Panel
Figure 9-39 Tune and Mass Calibration Status Menu
Mass . . . . . . . . . . . . . . . . . . . . . . . . The mass number of the peak.
Target Resolution. . . . . . . . . . . . . . Target Resolution at 10% peak height.
Actual Resolution. . . . . . . . . . . . . . Measured resolution at 10% peak height.
Resolution. . . . . . . . . . . . . . . . . . . . Resolution value; can be used to input a
change in Resolution value.
Ion Energy. . . . . . . . . . . . . . . . . . . . Ion Energy value; can be used to input a
change in Ion Energy value.
Position (DAC Value) . . . . . . . . . . . Current DAC setting for mass position.
Scan Width . . . . . . . . . . . . . . . . . . . Displays the points measured per amu.
Min OK Target Percentage . . . . . . Displays the minimum target percentage
required for the mass peak to meet the OK
LOW criteria.