8 - 44
HAPSITE ER Operating Manual
8.7.2.1 Additional Features of the Range Tool
By placing the mouse over the Range Tool triangle and right-clicking, the following
menu will be displayed. (See Figure 8-88.)
Figure 8-88 Range Tool Menu
Remove . . . . . . . . . . . . . . . . . . . . . . Removes the range cursor.
Select Mass for Integration . . . . . . Selects the TIC or a specific mass fragment
for reintegration.
Show Integration . . . . . . . . . . . . . . Displays the integration on the x-axis.
Show Retention Time. . . . . . . . . . . Displays the retention time on the x-axis.
Show Signal/Noise . . . . . . . . . . . . . Shows Signal to Noise ratio. A background
must be selected using Background
Subtract first. Refer to section 8.7.1,
Background Subtract, on page 8-38.
Subtract Current Spectrum . . . . . . Subtracts the spectrum at the point where
the green scan cursor is located from the
range.
8.8 Displaying Reconstructed Ion Chromatograms (RIC)
RIC plots are used to locate specific compounds in a chromatogram. A RIC plot of
the top three or more mass fragments can help locate the peak of interest. Follow
the instructions below to create a RIC plot.
NOTE: The NIST program uses the term peak instead of mass fragment.
However, the terms are synonymous.
Alternately, double-clicking on a mass in the Scan window will automatically
insert the selected mass in the Control Panel table and display the RIC for the
selected mass.
When the box in the Control Panel labeled -RIC is checked, the TIC/RIC window
will display the TIC minus the RIC selected.